Viso Light Inspector capacity list

Term Symbol Unit Definition Standards Viso Capacity Viso Data Extraction Keywords
Data presentation modes - All measurements are fundamentally radiometrical, and users can freely choose to have data presented in various modes
Photometric Measurement of light in terms of what is perceived by the human eye (default: absolute photometry. Relative photometry mode available). CIE 121, CIE S 025, LM-79, EN 13032-4 Default output in lumen (lm) and candela (cd)
Horticultural PAR photosynthetic active radiation PAR region (400-700 nm) output in PPF (mmol/s) and PPFD
Radiometric All physical radiation (in a given wavelength range) Radiated power output in W and W/sr
Dose units (radiometric) All physical radiation (in a given wavelength range combined with specific exposure input) Radiated power output in W and W/sr and specific exposure time (hh:mm)
Measurement ranges
Spectrometer ranges Viso standard software supports all measurement ranges. Viso BaseSensor/ LabSensor:
VIS 360 – 830 nm
UV-VIS 200 – 850 nm
UV-VIS-NIR 200 – 1100 nm
VIS-NIR 360 – 1100 nm
LightSpion VIS: 360 – 830 nm
Intensity range The intensity range is governed both by the sensitivity of the sensor and the measurement distance. For small and weak light sources a short sensor distance is recommended Viso BaseSpion - 25 mcd @0.35m to 2,000,000 cd @4.5m
Viso LabSpion- 0.2 cd @1m to 80,000,000 cd @20m
C-axis rotation range The Viso automatic c-plane resolution setting sets the necessary qty of c-planes automatically by starting the measurement with a pre-measurement that determines the assymmetry of the light distribution. Full manual control is also an option. 0°~+360°, automatic 4 to 72 c-planes, manual up to 360 c-planes
Gamma axis rotation range The Viso automatic gamma resolution setting determines the necessary resolution automatically. It starts out in 5° steps, and then reduces this resolution if necessary - but only in the beam section to save time. Full manual control is also an option. Viso BaseSpion and LabSpion = Automatic gamma rotation range: -180° to +180°, Manual resolution 0.1° to 5°. Automatic resolution 0.01° to 5°.
Viso LightSpion = Automatic gamma rotation range: -180° to +180°, Manual resolution 1° to 5°. Automatic resolution 1° to 5°.
Measurement range of luminous flux (Lambertian distribution) Viso BaseSpion - 0.08 lm @1m – 12,700,000 lm @10m max. +/- 4%
Viso LabSpion - 0.63 lm @1m – 63,000,000 @10m max. +/- 4%
Built-in AC power analyzer voltage range AC: The system includes an AC power outlet to the DUT. The system can be supplemented with a more advanced external AC power supply (Viso LabPower combined power supply and power analyser) and other AC power supplies and power analyser. DC: In parallel, the system can be supplemented with any DC supply, and USB communication can be established. 90 VAC - 260 VAC <+/- 0.5V
Built-in AC power analyzer current range 0 – 3 A (Avg: +/- 0.5 mA)
Built-in AC power analyzer power range 0 – 300 W (Avg: +/- 0.1 W)
Built-in AC power analyzer sample rate 70,000 samples/sec
Basic photometrical data
Luminous intensity I cd (= lm/sr) Density of luminous flux with respect to solid angle in a specified direction CIE S 017:2011 Yes {INT#} and {RINT#}
Luminous intensity distribution cd Presentation, by means of curves or tables, of the values of the luminous intensity of a source as a function of direction in space
Sample Image Sample Image Sample Image
Yes - tables, polar curves, linear distribution, 3D and iso-candela diagrams. Output as IES files, LDT files, TM33 (XML) files
Sample Image
Luminous flux Φ lm Luminous flux is the energy measurement of visible light treleased by a source of light. CIE S 017:2011 Yes {LUM#}
Relative photometry efficiency η % The difference between the measured light source lumen and the lumen of the naked light source used in the lighting fixture. Yes {REL_EFF} and {SOURCE_LUM}
Beam angle degrees (°) Angle between two imaginary lines in a plane through the optical beam axis, such that these lines pass through the centre of the front face of the device and through points at which the luminous intensity is 50% of the centre beam intensity Yes {B_ANG#}
Field angle (beam spread) degree (°) The angle between the two directions for which the intensity is 10% of the maximum intensity as measured in a plane through the nominal beam centerline Yes {F_ANG#}
Cutoff angle degrees (°) The angle between the two directions for which the intensity is 2,5% of the maximum intensity as measured in a plane through the nominal beam centerline yes {C_ANG#}
Uplight / Downlight % Luminous flux in the upper/lower hemisphere + up/down flux fractions Yes {LUM_UP} / {LUM_DOWN}
{UFF} / {DFF}
Illuminance distribution E lux (=lm/m2)
fc (=lm/ft2)
Presentation, by means of curves or tables, of the density of incident luminous flux with respect to area at a point on a real or imaginary surface
Sample Image Sample Image
Yes - tables and iso-lux/footcandle diagrams
Half illuminance angle(s) degrees (°) The angle where the illuminance is half the peak illuminance on a plane perperdicular to the gamma 0 direction {HALFILLUMANG} average, {HALFILLUMANGC0} plane c0-180 , {HALFILLUMANGC90} plane c90-270
Beam distance m The distance in front of the lamp where is illuminance has dropped to 0.25 lux Yes {BEAMDISTL_M}
Beam width m or ft Beam width in m or ft from the lamp Yes {BW_M#} / {BW_F#}
Peak plane degrees (°) The c-plane that contains the maximum intensity (cd) measured Yes {PEAK} and {PEAKPLANETEXT}
BUG rating Backlight-Uplight-Glare rating IESNA TM-15-11 Yes - BUG graphics and data
Sample Image
{BUG#}
Derived photometrical data
UGR tables Unified Glare Rating (UGR) CIE 117:1995 (Uncorrected, comprehensive, S/H=0.25) Yes
Utilization factors Quotient of the luminous flux received by a reference surface and the sum of the rated individual luminous fluxes of the lamps of the installation Yes - special utilization factor graphics, also JP version
Sample Image
Luminaire budgetary diagram Budget for qty of luminaires as a function of indoor space area and reflectance values. Yes - special luminaire budgetary diagram
Sample Image
NITS (for uniform luminance distributions only) nt (= 1 cd/m2) Nits are a unit of measurement used to quantify the brightness of electronic displays, particularly those with emissive technologies like LEDs, OLEDs, and microLEDs. One nit is equivalent to one candela per square meter (cd/m²) Yes
EU Ecodesign directionality Yes - returns DLS = Directional Light Source and NDLS = Non Directional Light Source {EU_DIR}
CIE Flux code CIE Luminaire classification system EN 13032-2 AND CIE 040-1978 Yes - via custom PDF reports
EN G* class EU outdoor glare classification system EN 13201-2 G* class Yes - via custom PDF reports
Dark sky compliance Test for US dark sky association luminaire compliance Yes - via custom PDF reports
EU Ecodesign Useful Luminous Flux lm Yes {EU_USEFUL}
Derived human circadian light response data
Melanopic EDI lux Melanopic Equivalent Daylight Illuminance CIE 026 a-opic Toolbox - v.1.049a - 2020/11 Yes - for a given angle and distance {MEDILX#} (and {PHOLX#})
MP Ratios M/P 1 - M/P 2 - M/P 3 (Standard WELL v2-2019) - M/P 4 (Recommended by CIE S026-2018) CIE S 026:2018 Yes - via custom PDF reports
Melanopic action factor The melanopic action factor is the ratio of melanopic irradiance to photopic irradiance CIE S 026:2018 Yes - for a given angle and distance {MELA_ACTION_FACTOR#}
Human retinal photopigment complement Melanopic (melanopsin), Rhodopic (rods), Cyanopic (s cone), Chloropic (M cone), Erythropic (L cone) Yes - via custom PDF reports
Special spectral response data E.g., scotopic light, S/P ratios, mesopic light,
COI Cyanosis observation index AS/NZS 1680.2.5 Yes - via custom PDF reports
Spectral G-index Protection for astronomical or ecological reasons Yes - via custom PDF reports
Basic radiometric data
Radiant intensity I W/sr Density of radiant flux with respect to solid angle in a specified direction Yes
Peak irradiance (@ 1 m)
Radiant Intensity distribution Presentation, by means of curves or tables, of the values of the radiant intensity of a source as a function of direction in space Yes
Radiant flux P W Radiant flux is the energy measurement of total electromagnetic energy treleased by a source of light. Yes
Basic horticultural data
PPF - photosynthetic photon flux 400-700 nm µmol/s Total photon flux Yes {PPF#}
PPFD - photosynthetic photon flux density 400-700 µmol/(m2·s) Peak photon irradiance at peak at one meter distance. PPFD is measured in micromoles per square metre per second (µmol/m2/S) which establishes exactly how many PAR photons are landing on a specific area Yes {PPFD}, {PPFD_F#} (specific distance in ft), {PPFD_M#} (specific distance in m)
PPFW - photon efficiency µmol/J Photon efficiency in µmol per Joule Yes {PPFW}
DLI - Daily light integral µmol/(m2·day) DLI measures the sum of photosynthetic light (light in PAR range, 400-700nm) that plants receive over the course of a full day (a 24-hour period) Yes {DLI}
Photosynthetic photon flux for the whole spectrum µmol/s Total photon flux for the integrated spectrum (or a specific range) Yes {PFLUX#}
Color data
Spectral power distribution (SPD) W/nm Radiated power per wavelength Yes - per measurement point or full light source (=integrated spectrum) - including graphics
Sample Image
{PNM} and {PERNM} (the percentage over light in a given wave
Peak(s) in spectral power distribution nm Wavelength per peak no. Yes - numbered peaks {PEAK#}
Dominant wavelength nm Yes {DNM}
Correlated color temperature (CCT) T K Temperature of a Planckian radiator having the chromaticity nearest the chromaticity associated with the given spectral distribution on a modified 1976 UCS diagram where u′, are the coordinates of the Planckian locus and the test stimulus Yes - per measurement point or full light source (=integrated CCT) {CCT} and {RCCT}
Color Rendering Index (CRI) Measure of the degree to which the psychophysical colour of an object illuminated by the test illuminant conforms to that of the same object illuminated by the reference illuminant CIE 13 Yes - Specific values avaible too (R1 through R14) and graphics
Sample Image
{R} and {R#} (14 specific values)
CQS (Color Quality Scale) NIST Color quality system Yes - including graphics
Sample Image
{Q#}
Television lighting consistency index 2012 TLCI A single averaged score up to 100 that is determined by the color response that would result when using a video camera. ANSI/IES TM-30-18 Yes - including graphics
Spectral Similarity Index SSI Spectral Similarity Index (SSI), Calculated according to the Society of Motion Picture and Television Engineers: SMPTE standard ST 2122:2020 – Spectral Similarity Index (SSI)” Yes - via special PDF report
ANSI/IES TM-30-18 Color Fidelity Rf Color Fidelity: Enumerates the accuracy with which the color appearance of surfaces and objects under a given test source match their appearance under a reference illuminant. ANSI/IES TM-30-18 Yes - including graphics
Sample Image
{TM3} and {C#} (99 specific values)
ANSI/IES TM-30-18 Color Gamut Rg Color Gamut: Enumerates the average increase or decrease of the chroma of surfaces and objects when viewed under a given test source relative to when viewed under a reference illuminant. ANSI/IES TM-30-18 Yes {TM3g}
ANSI/IES TM-30-18 Color Graphics
Sample Image

Sample Image

Sample Image
ANSI/IES TM-30-18 Yes - Color vector graphics, color distortion graphics, Rf by Hue, Chroma shift by hue, and Rf/Rg plot
ANSI/IES TM-30-18 Special output ANSI/IES TM-30-18 Yes {TM3fh1} Rf,h1 value; {TM3csh1} Rcs,h1 value; {TM3P} Design intent Preference; {TM3V} Design intent vividness, {TM3F} Design intent fidelity
MacAdam Ellipses SDCM ANSI C78-376 Yes - including ANSI C78-376 Ellipses, and CIExy graphics
Sample Image
{SDCM}, {SDCM_ANSI}
Color deviation from BBL CIE TN 001:2014 Yes - via special PDF report
Duv (chromaticity metric) Closest distance from the Planckian locus on the (u', 2/3 v') diagram, with + sign for above and - sign for below the Planckian locus. ANSI C78.377-2008 Yes - per measurement point or full light source {DUV} and {DuvMarkLinDist#}
CIE 1931 Color coordinates x,y Yes - including graphics
Sample Image
{CIEx#}, {CIEy#}, {RCIEx#}, {RCIEy#}
CIE 1960 Color coordinates u,v CIE ISO 11664-5:2024 Yes {CIEu#}, {CIEv#}, {RCIEu#}, {RCIEv#}
CIE 1976 Color coordinates u',v' CIE 015:2004 Yes {CIEuMARK#}, {CIEvMARK#}, {RCIEuMARK#}, {RCIEvMARK#}
Power and efficacy data (obtained with built-in power analyzer, sampling rate 70,000 samples/s)
Power consumption P Yes
Range 0 – 300 W (Avg: +/- 0.1 W)
{PWR}
Voltage U V Yes
Range 90 VAC - 260 VAC max. +/- 0.5V
{VOL}
Current I A Yes
Range 0 – 3 A (Avg: +/- 0.5 mA)
{CUR}
Frequency Hz Yes {FREQ}
Power factor Yes {PF}
Displacement factor Yes {DPF}
Luminous efficacy lm/W CIE S 017:2011 Yes {LPW}
Radiant efficacy W/W CIE S 017:2011 {LPW}
Total harmonic distortion For current and voltage
Sample Image
Yes- including graphics
Sample Image
{THDC} and {THDV}
EU Ecodesign energy class Yes - Returns A,B,C, etc {EU_ECLASS}
Power and efficacy data (obtained with Viso LabAnalyzer, sample rate 2,000,000 samples/s)
Power consumption P CIE S 025/ IES LM-79 /EN 13032-4 Yes
Power range AC 0.05W to 1700W @ 240VAC / 850W @ 120 VAC
Power range DC 0.05W to 1700W @ 240VDC / 850W @ 120 VDC / 170W @ 24VDC / 85W @ 12VDC
(50/60 Hz +/- 0.5%)
{PWR}
Voltage U V CIE S 025/ IES LM-79 /EN 13032-4 Yes
Range 0 - 826 V max. +/- 0.2% of reading+0.2% of range
{VOL}
Current I A CIE S 025/ IES LM-79 /EN 13032-4 Yes
Range 3 mA – 16.5 A (max. 0.3% of reading + 0.3% of range)
{CUR}
Frequency Hz Yes
Range 0-100 kHz
{FREQ}
Power factor Yes {PF}
Displacement factor Yes {DPF}
Flicker data (obtained with the Viso LabFlicker accessory)
Percent flicker = Modulation depth CIE TN 006:2016. Yes {FLP}
Flicker Index CIE S 017:2020, 17-22-094 (CIE 2020) Yes {FLI}
Flicker frequency Yes {FLF}
SVM (Stroboscopic visibility measure) MVS CIE TN 006:2016 and IEC TR 63158:2018 Yes {FLS}
PstLM (visible flicker) Pst IEC TR 61547-1:2020 Yes {FLPSTLM}
Flicker Mp value A flicker perception metric recommended by ASSIST (Alliance for Solid-State Illumination Systems and Technologies, US) ASSIST vol. 11, Issue 3, 2015 Yes {FLMP}
Flicker sample rate F Hz Yes {FL_SAMPLE_RATE}
JA8/JA10 Test method described by California energy commission in title 24: Percent amplitude modulation at 40 Hz, 90 Hz, 200 Hz, 400 HZ, and 1000 Hz California Energy Commission (CEC) 2016b Yes {FL_CO40}, {FL_CO90}, {FL_CO200}, {FL_CO400}, {FL_CO1000}
Flicker waveform and harmonics Sample Image
Sample Image
Yes - including graphics
IEEE 1789 frequency modulation plot Sample Image Yes - via special PDF reports